The article “The Noise Power Spectrum of CT Images” written by Marie Foley Kijewski and Philip F. Judy, published on May 1, 1987, makes a significant contribution to the analysis of the noise power spectrum (NPS) in computed tomography (CT) images reconstructed using the discrete filtered backprojection algorithm. The authors derive an expression for the NPS that explicitly includes sampling within the projections, angular sampling, and the two-dimensional sampling implicit in the discrete representation of the images. They also consider the effects of interpolation.
The article explores how the NPS predictions based on their analysis differ from continuous theories in two crucial aspects: the rotational asymmetry of the NPS and the presence of components at low frequencies that do not tend to zero. These features are attributed to two-dimensional aliasing due to pixel sampling. The predictions were confirmed by measuring the NPS in simulated images and in images obtained from a commercial X-ray transmission CT scanner.
The initial analysis focuses on considering various sampling processes during image reconstruction. The noise in the raw projection data is assumed to be stationary, additive, and uncorrelated. The projection NPS is constant and limited to frequencies below the Nyquist frequency determined by the projection sampling distance.
In the backprojection process, the points in the projections where data are required for reconstruction generally do not coincide with the points where the data were measured. This requires interpolation, which converts the discrete projection into a continuous one, whose power spectrum, however, still exhibits replicas due to the initial discretization.
The article also addresses how convolution filters, such as the apodized ramp filter, affect the NPS. Specifically, it highlights the importance of interpolation during backprojection and how such interpolation significantly modifies the expected NPS. The most common interpolations, such as nearest neighbor and linear interpolation, cause notable differences in the apodization of the spectrum; linear interpolation, in particular, attenuates the high-frequency components, which manifests as an “apodization” in the NPS analysis.
Experimental results include NPS from both simulated and real images obtained with a commercial CT scanner (Technicare 2020), reconstructed using ramp filters and different interpolation methods. The authors discuss how their findings differ markedly from previous analyses, such as those conducted by Faulkner and Moores (1984), who did not consider the two-dimensional effects of sampling and interpolation on the NPS.
In conclusion, this study reveals that the two-dimensional aliasing resulting from pixel sampling destroys the rotational symmetry of the NPS of CT images and that aberrations in low-frequency components and the zero-frequency (DC) component can have significant implications for detectability and accurate quantification in CT. These insights offer a more comprehensive and precise understanding of noise behavior in CT images, crucial for improving image processing and reconstruction techniques in medical applications.