A comparison between Epson V700 and Epson V800 scanners for film dosimetry

0
408

The article titled “A comparison between EPSON V700 and EPSON V800 scanners for film dosimetry” by Shameem Tarafder, Nick Bennie, Martin Butson, and David Thwaites, published in March 2020, explores the differences between two EPSON scanner models, V700 and V800, in the context of radiochromic film dosimetry for complex radiotherapeutic treatment techniques. This study is essential as commercial EPSON scanners are frequently used in film dosimetry due to their high spatial resolution and the near equivalence of radiochromic films to human tissues.

The research starts from the premise that the EPSON V800 is an upgrade of the V700 model, but without clear manufacturer guidelines on their differences beyond the fact that the V800 uses a white LED light source instead of the cold fluorescent tube of the V700. Therefore, it raises the need to evaluate whether these two models can be used interchangeably in film dosimetry.

Tests were conducted using EBT3 films irradiated with 6 MV beams and a field size of 40×40 cm², scanned on both the V700 and the V800. The scanned data were processed with ImageJ software and compared in MS Excel. Among the evaluated aspects are the lateral response artifact (LRA) effect, light polarization, scanner response, repeated scan effects, calibration curves, noise, and scanner warm-up time.

One of the main conclusions of the study is that the V800 introduces more light polarization, which significantly increases the LRA effect compared to the V700. The V800 scanner also shows variations in the color channel responses, especially in the red channel, where the response is up to 26.9% higher than in the V700, while in the blue and green channels, the differences are 2.0% and 6.2%, respectively.

Additionally, it was found that the V700 shows a notable darkening effect after repeated scans, whereas in the V800 this effect is less pronounced, which is attributed to the different light sources used in the scanners. In terms of noise, the V800 was noisier in the blue and green channels, while the V700 was noisier in the red channel. Similarly, the calibration curves of both scanners show small differences at low doses (up to 3 Gy), but become more significantly distinct at higher doses, particularly in the red channel.

The article concludes that, due to differences in scanner response, LRA effect, noise, and calibration curves, it is not possible to use the V700 scanner interchangeably with the V800 without proper characterization and commissioning, as well as correction of the specific factors for each model. The study emphasizes the need for a thorough evaluation before implementing an alternative scanner model in clinical practice to ensure accuracy and precision in radiochromic film dosimetry.